Technical Data
Power Supply: 100 – 240 V AC, 50/60 Hz
Battery: Lithium-Ion battery, rechargeable, 10.8 V DC, 6,600 mAh, Charging time to a
100%: 6 hours, Operating time approx. 11 hours (new and fully charged battery)
Monitor Dimensions: L x H x W: 303 x 230 x 110 mm (11.93 x 9.05 x 4.33 in)
Weight: 3600 g, (7.94 lb)
Basket Dimensions: L x H x W: 270 x 225 x 200 mm (10.63 x 8.86 x 7.87 in)
Alarm: Visible LEDs (Yellow, Red, Blue, Green, Orange)
Audible Speaker
Gives audible alarm, QRS tone
Ports: 1 x AC power inlet
1 x standard RJ45 interfaces.100 BASE-TX, IEEE 802.3
4 x USB ports
1 x Equipotential grounding point
1 x RJ11 connector for nurse call
DC out port 15V/1.2 A
Clinical-grade NIBP (non-invasive blood pressure)
Measuring method: automatic oscillometric
Measurement range: Adult
SYS 30 – 270 mmHg, DIA 10 – 220 mmHg, MAP 20 – 235 mmHg
Pediatric
SYS 30 – 235 mmHg, DIA 10 – 220 mmHg, MAP 20 – 225 mmHg
Neonate
SYS 30-135 mmHg, DIA 10-110 mmHg, MAP 20-125 mmHg
Cuff pressure range: 0 – 280 mmHg
Resolution: 1 mmHg
Pressure accuracy: Static ±3 mmHg
Clinic ±5 mmHg, standard deviation _ 8 mmHg
Pulse rate range: 40 – 240 bpm
Pulse rate accuracy: 1 bpm
Initial inflation pressure: Adult 160 mmHg, Pediatric 130 mmHg, Neonate 75 mmHg
Double overpressure protection (hardware and software)
Adult 297 ± 3 mmHg, Pediatric 252 ± 3 mmHg , Neonate 147 ± 3 mmHg
Alarm range:
SYS 0 – 300 mmHg
DIA 0 – 300 mmHg
MAP 0 – 300 mmHg
Riester SpO2 Measurement range: 0 – 100%
Resolution: 1%
Accuracy: ±2%(70 to 100%)
±3% (40% to 69%)
Unspecified (0% to 39%)
Alarm range: 0 – 100%
Average time: 4 sec, 8 sec, 16 sec
Pulse rate
Measurement range: 20 – 250 bpm
Resolution: 1 bpm
Accuracy: ±1% or ±1 bpm whichever is greater
Nellcor SpO2 Measurement range: 0-100%
Resolution: 1%
Accuracy: 70 – 100%: ±2% (adult, pediatric)
70 – 100%: ±3% (neonate)
0 – 69%: unspecified
Alarm range: 0 – 100%
Average time: 8 sec, 16 sec
Pulse rate
Measurement range: 20 – 300 bpm
Resolution: 1 bpm
Accuracy: 20 – 250 bpm: ±3 bpm
251 – 300 bpm: unspecified
Masimo SpO2 Measurement range: 0 – 100%
Resolution: 1%
Accuracy: 70 – 100%: ±2% (adult, pediatric, non-motion conditions)
70 – 100%: ±3% (neonate, non-motion conditions)
70 – 100%: ±3% (motion conditions)
0 – 69%: unspecified
Average time: 2 – 4 sec, 4 – 6 sec, 8 sec, 10 sec, 12 sec, 14 sec, 16 sec
Pulse rate
Measurement range: 25 – 240 bpm
Resolution: 1 bpm
Accuracy: ±3 bpm (non-motion conditions)
±5 bpm (motion conditions)
Predictive thermometer
- Oral ( Quick Mode ) non-fever temps: 3-5 seconds
fever temps: 8-10 seconds
- Oral ( Standard Mode ): 6-10 seconds
- Axillary Mode: 8-12 seconds
- Rectal Mode: 10-14 seconds
- Direct Mode (all sites): 60-120 seconds